Programmable Constant Temperature and Humidity Testing Machine(touch screen)
Product Details
Constant temperature and humidity testing machine is also called environment testing machine, used to test various types of materials for their tolerances of heat, cold, dry, humidity.Application: Quality inspection for all types of electronic products, electric appliance, and products from communication, instrument, automobile, plastic, metal, food, chemical, building materials, medical and aerospace industries.
Features
- Graceful appearance, circularly shaped body, surface treated with mist strips, and plane handle with no reaction. easy to operate, safe and reliable.
- Rectangular double-glassed watching window for the observation of the test products during the testing process. the window is equipped with sweat-proof electrical heating device that can prevent water steam from condensing into droplets, and with high brightness PL fluorescent bulbs to provide light inside the box.
- Double-layer-insulated airtight doors, able to insulate the internal temperature effectively
- Water supply system that is externally connectable, convenient for refilling water into the humidifying pot and automatically recyclable.
- The France¡¯s "Taikang" brand is used for the circulation system of the compressor, capable of removing the lubricant between the condensation pipes and capillaries. The environment-protecting coolant is used for the whole series (R23/R404/R507).
- Imported LCD display screen is used for the control unit, capable of displaying the measured value as well as the set value and time.
- The control unit has the functions of multiple segment program editing, and of quick or slope control of temperature and humidity.
- Inserted mobile pulley, convenient for movement and relocation, with strong positioning screws.
Implementation standards and methods
- GB11158 temperature Chambers technical conditions
- GB10589-89 Chambers cryogenic technology
- GB10592-89 Chambers technology high and low temperature conditions
- Chambers technology GB/T10586-89 hot and humid conditions
- GB/T2423.1-2001 hypothermia Chambers Test Method
- High Temperature Test Method Chambers GB/T2423.2-2001
- GB/T2423.3-93 hot and humid Chambers Test Method
- Alternating hot and humid GB/T2423.4-93 Test Method
- GB/T2423.22-2001 temperature changes Test Method
- IEC60068-2-1.1990 low temperature Chambers Test Method
- IEC60068-2-2.1974 high temperature Chambers Test Method
- GJB150.3 High Temperature Test
- GJB150.4 Low Temperature Test
- GJB150.9 Damp Heat Test
SPEC
| Model | CZ-A-80 (A~G) | CZ-A-150 (A~G) | CZ-A-225 (A~G) | CZ-A-408 (A~G) | CZ-A-608(A~G) | CZ-A-800 (A~G) | CZ-A-1000 (A~G) |
| Internal Dimensions WxHxD (cm) | 40x50x40 | 50x60x50 | 50x75x60 | 60x85x80 | 80x95x80 | 100x100x80 | 100x100x100 |
| External Dimensions WxHxD (cm) | 90x136x94 | 100x146x104 | 100x161x117 | 110x171x137 | 130x181x137 | 150x186x137 | 150x186x157 |
| Temperature Range | -70℃~+100℃(150℃) (A:+25℃ B:0℃ C:-20℃ D:-40℃ E:-50℃ F:-60℃ G:-70℃) | ||||||
| Humidity Range | 20%~98%R.H.(10%-98%R.H/5%~98%R.H is a special optional condition) | ||||||
| Analytic Accuracy/evenness degree of Temperature and Humidity |
±0.1℃;±0.1%R.H./2.0℃;±3.0%R.H. | ||||||
| Control Accuracy/Temperature stabicity of Temperature and Humidity | ±0.3℃;±2.5%R.H. | ||||||
| Temperature Rising/Falling Time | Approximately 4.0℃/minute; approximately1.0℃/minute ("5~10℃ decrease per minute" is a special optional condition) | ||||||
| Inner and Outer Materials | The whole set is SUS 304# stainless steel treated with mist. Inner box is stainless steel | ||||||
| Insulation Materia | Resistant to high temperature, high density, formate chlorine, ethyl acetum foam insulation materials | ||||||
| Cooling System | Air cooling /single segment compressor (-20℃), air water cooling/double segment compressor | ||||||
| Protection Devices | Fuse-free switch, overloading protection switch for compressor, high and low voltage coolant protection switch, over-humidity and over-temperature protection switch, fuses, failure warning system | ||||||
| parts | Recorder (optional), watching window, 50mm testing hole, PL internal bulbs,partition plate, dry and wet ball gauze | ||||||
| control panel | South Korea"TEMI"or Japan¡äs "OYO" Brand ¡¡Optional | ||||||
| Compressor | France¡äs "Tecumseh" Brand | ||||||
| Power | 220VAC±10% 50/60Hz 380VAC±10% 50/60Hz | ||||||
Relevant
- Temperature and Humidity Testing Machinge
- Cold and thermal shock Testing Machine
- High And Low Temperature Testing Machine
- Low Temperature (cold) Testing Machine
- Hot Circulator Exact Oven
- Saline Spray Testing Machine
- Aging test machine
- Temperature/humidity/ vibration Integrated Environmental Tester System

400){this.resized=true;this.style.width=400;}¡ä alt="Programmable Constant Temperature and Humidity Testing Machine(touch screen)">